Apply Corrections - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

deletion is only performed if you press OK or Enter.
Key Path
Remote Command
Example
Notes
Initial S/W Revision

Apply Corrections

Applies amplitude corrections, which are marked as ON to the measured data. If this is set to OFF, then no
amplitude correction sets will be used, regardless of their individual on/off settings. If set to ON, the
corrections that are marked as ON (see
Key Path
Remote Command
Example
Preset
State Saved
Initial S/W Revision
Delete All Corrections
Erases all correction values for all 4 Amplitude Correction sets.
When this key is pressed a prompt is placed on the screen that says "Please press Enter or OK key to
delete all corrections.  Press ESC or Cancel to close this dialog."  The deletion is only performed if you
press OK or Enter.
Key Path
Remote Command
Example
Initial S/W Revision
EMI Receiver Mode Reference
Input/Output, Corrections
[:SENSe]:CORRection:CSET[1]|2|...|6:DELete
CORR:CSET:DEL
CORR:CSET1:DEL
CORR:CSET4:DEL
Pressing this key when no corrections are present is accepted without error.
A.02.00
"Correction On/Off" on page
Input/Output, Corrections
[:SENSe]:CORRection:CSET:ALL[:STATe] ON|OFF|1|0
[:SENSe]:CORRection:CSET:ALL[:STATe]?
SENS:CORR:CSET:ALL OFF
This command makes sure that no amplitude corrections are applied, regardless of their individual
on/off settings.
Not affected by Preset. Set to OFF by Restore Input/Output Defaults
Saved in instrument state.
A.02.00
Input/Output, Corrections
[:SENSe]:CORRection:CSET:ALL:DELete
CORR:CSET:ALL:DEL
A.02.00
4  Input/Output Functions
184) are used.
Input/Output
193