8  Frequency Scan Measurement
Meas Setup
Preset
State Saved
Min
Max
Default Unit
Backwards
Compatibility SCPI
Backwards Compatibility
SCPI Notes
Initial S/W Revision

Scan Time

Specifies the scan time of the selected range.
Key Path
Remote Command
Example
460
No rules for entries during Discrete (Stepped) Scan.
If user changes the step size to a value that causing the total range point to exceed 40001, range
state of the specific range will be changed to off. An advisory message is generated. If the equivalent
SCPI command is sent, this same message is generated as part of a "–221, Settings conflict"
warning.
Auto Step Size and Auto Scan Points is always coupled.
100 Hz|4.5 kHz|60 kHz|60.003 kHz|60.003 kHz|500kHz|100 kHz|100 kHz|100 kHz|100 kHz
ON
Saved in instrument state.
The minimum step size will be limited to a value where the number of points for a range cannot be
more than 40001.
Depends on the maximum range of the selected range, but the value is limited by the Max frequency
of the instrument.
Option 503 (3 GHz models): 3.08 GHz
Option 503 (3.6 GHz models): 3.7 GHz
Option 507 (7 GHz models): 7.1 GHz
Option 507 (7.5 GHz models): 7.58 GHz
Option 508: 8.5 GHz
Option 513: 13.8 GHz
Option 526: 27.0 GHz
Hz
:SENSe1|SENSe2:SCAN[1]|2|...|10:STEP
[:SENSe]:SCAN[1]|2|...|10:STEP
This command above is included for ESU compatibility. The suffix 1|2 is irrelevant to the
measurement, they are added solely for remote language compatibility.
A.07.00
Meas Setup, Scan Table
[:SENSe]:FSCan:SCAN[1]|2|...|10:STIMe <time>
[:SENSe]:FSCan:SCAN[1]|2|...|10:STIMe?
[:SENSe]:FSCan:SCAN[1]|2|...|10:STIMe:AUTO ON|OFF|1|0
[:SENSe]:FSCan:SCAN[1]|2|...|10:STIMe:AUTO?
FSC:SCAN7:STIM 0.5  Set the dwell time of Range 7.
FSC:SCAN7:STIM?
FSC:SCAN:STIM:AUTO ON ! Turn on Auto Scan Time in Range 1.
FSC:SCAN:STIM:AUTO?