Marker Trace - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

11  APD Measurement
Marker
Key Path
Remote Command
Example
Notes
Preset
State Saved
Min
Max
Initial S/W Revision

Marker Trace

This assigns the specified marker to the designated trace.
Key Path
Remote Command
Example
Preset
State Saved
Range
Initial S/W Revision
Couple Markers
When this function is true, moving any marker causes an equal X axis movement of every other marker
which is not Off. By "equal X axis movement" we mean that we preserve the difference between each
marker's X axis value (in the fundamental x-axis units of the trace that marker is on) and the X axis value of
the marker being moved (in the same fundamental x-axis units).
Key Path
Remote Command
Example
922
Marker, Properties
:CALCulate:APD:MARKer[1]|2|...|12:REFerence <integer>
:CALCulate:APD:MARKer[1]|2|...|12:REFerence?
CALC:APD:MARK1:REF 3
CALC:APD:MARK2:REF?
A marker cannot be relative to itself so that choice is grayed out, and if sent from SCPI generates
error –221: "Settings conflict; marker cannot be relative to itself."
When queried a single value will be returned (the specified marker numbers relative marker).
2|3|4|5|6|7|8|9|10|11|12|1
Saved in instrument state
1
12
A.13.00
Marker, Properties
:CALCulate:APD:MARKer[1]|2|...|12:TRACe MEAS|MEAS2|ELIMit
:CALCulate:APD:MARKer[1]|2|...|12:TRACe?
CALC:APD:MARK3:TRAC MEAS
CALC:APD:MARK2:TRAC?
Measured| Measured 2| Exceeded Limit
Saved in instrument state
Measured
A.13.00
Marker
:CALCulate:APD:MARKer:COUPle[:STATe] OFF|ON|0|1
:CALCulate:APD:MARKer:COUPle[:STATe]?
CALC:APD:MARK:COUP ON  Sets Couple Markers on.
EMI Receiver Mode Reference

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pxa n9030aMxa n9020aExa n9010aCxa n9000aEmi n9038a

Table of Contents