Keysight x-series User Reference Manual page 317

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

Remote Command
Example
Notes
Dependencies
Couplings
Initial S/W Revision
Parameter Name
Mode
Remote Command
Example
Notes
Dependencies
Couplings
Force Restart
Initial S/W Revision
Parameter Name
Mode
Remote Command
Example
Notes
Dependencies
Couplings
Force Restart
Initial S/W Revision
EMI Receiver Mode Reference
:CALibration:REFerence:CLOCk?
:CAL:REF:CLOC:INIT?
//connect cable
:CAL:REF:CLOC?
//disconnect cable
:CAL:REF:CLOC:END?
:CALibration:REFerence:CLOCk? returns 0 if successful
:CALibration:REFerence:CLOCk? returns 1 if failed
Option DP2 or B40
Initializes the time for the Last Characterize Reference Clock Time.
Records the temperature for the Last Characterize Reference Clock Temperature. Expected to be run
after :CAL:REF:CLOC:INIT, and before :CAL:REF:CLOC:END.
A.13.00
Characterize Reference Clock Initialization
All
:CALibration:REFerence:CLOCk:INITialize?
:CAL:REF:CLOC:INIT?
:CALibration:REFerence:CLOCk:INIT? returns 0 if successful
:CALibration:REFerence:CLOCk:INIT? returns 1 if failed
Option DP2 or B40
Expected to be run before sending the :CAL:REF:CLOC? command.  This will stop the current
measurement when it has completed (does not abort the current data acquisition), and it will prepare
the instrument for the expected cabling.
Yes
A.12.00
Characterize Reference Clock End
All
:CALibration:REFerence:CLOCk:END?
:CAL:REF:CLOC:END?
:CALibration:REFerence:CLOCk:END? returns 0 if successful
:CALibration:REFerence:CLOCk:END? returns 1 if failed
Option DP2 or B40
Expected to be run after sending the :CAL:REF:CLOC? command, and after removing the cable used
in that Characterize Reference Clock step.  This will resume any queued measurements, and it
concludes the reference clock characterization.
Yes
A.12.00