10  Monitor Spectrum Measurement
Marker
Example
Notes
Preset
State Saved
Min
Max
Initial S/W Revision

Marker Y Axis Value ( Remote Command Only )

Returns the marker Y Axis value in the current marker Y Axis unit.
Remote Command
Example
Notes
State Saved
Initial S/W Revision
764
CALC:MON:MARK3:X 0
CALC:MON:MARK3:X?
If no suffix is sent, it will use the fundamental units for the current marker X Axis Scale.  If a suffix is
sent that does not match the current marker X Axis Scale unit, an "Invalid suffix" error will be
generated. 
The query returns the marker's absolute X Axis value if the control mode is Normal, or the offset from
the marker's reference marker, if the control mode is Delta.   The query is returned in the
fundamental units for the current marker X Axis scale: seconds.  If the marker is off the response is
not a number (NAN).
0
No
Current start frequency
Current stop frequency
A.13.00
:CALCulate:MONitor:MARKer[1]|2|...|12:Y?
CALC:MON:MARK11:Y?
The query returns the marker Y-axis result.  If the marker is Off the response is not a number (NAN).
No
A.13.00
EMI Receiver Mode Reference