Alignment Data Wizard - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

alignment that pertains to the instrument must be transferred to the resident hard drive after a CPU or
hard drive is replaced.  This utility facilitates backing-up and restoring the alignment data.
This utility allows the operator to navigate to any location of the Windows file system.  It is intended that
the operator use a USB memory device or Mapped Network Drive to back up the alignment data to storage
outside of the instrument.
Key Path
Initial S/W Revision
Key Path
Mode
Remote Command
Example
Couplings
Initial S/W Revision

Alignment Data Wizard

The Backup or Restore Alignment Data wizard guides you through the operation of backing-up or restoring
the alignment data.
The following dialogue boxes  operates without a mouse or external keyboard when you use the default file
names.
EMI Receiver Mode Reference
System, Alignments
A.02.00
System, Alignments
All
:CALibration:DATA:DEFault
:CAL:DATA:DEF
Sets Auto Align to Off. Sets bit 14 in the Status Questionable Calibration register. The Error Condition
message "Align Now, All required" is generated.
Prior to A.02.00
6  System Functions
System
309