alignment that pertains to the instrument must be transferred to the resident hard drive after a CPU or
hard drive is replaced. This utility facilitates backing-up and restoring the alignment data.
This utility allows the operator to navigate to any location of the Windows file system. It is intended that
the operator use a USB memory device or Mapped Network Drive to back up the alignment data to storage
outside of the instrument.
Key Path
Initial S/W Revision
Key Path
Mode
Remote Command
Example
Couplings
Initial S/W Revision
Alignment Data Wizard
The Backup or Restore Alignment Data wizard guides you through the operation of backing-up or restoring
the alignment data.
The following dialogue boxes operates without a mouse or external keyboard when you use the default file
names.
EMI Receiver Mode Reference
System, Alignments
A.02.00
System, Alignments
All
:CALibration:DATA:DEFault
:CAL:DATA:DEF
Sets Auto Align to Off. Sets bit 14 in the Status Questionable Calibration register. The Error Condition
message "Align Now, All required" is generated.
Prior to A.02.00