10 Monitor Spectrum Measurement
Marker
Normal Marker State (Remote Command Only)
Toggles the Normal marker state on and off.
Remote Command
Example
Notes
Preset
Initial S/W Revision
Normal Marker X Axis Value (Remote Command Only)
Sets the selected marker to Normal mode, and sets/ gets the marker X Axis Values.
Remote Command
Example
Notes
Preset
Initial S/W Revision
Normal Marker Y Axis Value (Remote Command Only)
Sets the selected marker to Normal mode, and returns the marker Y Axis Values.
Remote Command
Example
Notes
Initial S/W Revision
Delta Marker State (Remote Command Only)
Toggles the Delta marker state on and off.
Remote Command
Example
762
:CALCulate[1]|2:MARKer[1]|2|...|4[:STATe] ON|OFF|1|0
:CALCulate[1]|2:MARKer[1]|2|...|4[:STATe]?
CALC:MARK4 ON Turn on maker 4 in Normal mode.
This SCPI is Meas Local and Context Sensitive. This SCPI alias to :CALC:MON:MARK[1]|2|3|4:MODE
POS.
OFF
A.13.00
:CALCulate[1]|2:MARKer[1]|2|...|4:X <value>
:CALCulate[1]|2:MARKer[1]|2|...|4:X?
CALC:MARK3:X 3e4 Turn on maker 3 in Normal mode, set the marker frequency to 30 kHz.
CALC:MARK3:X? Turn on maker 3 in Normal mode and query the X axis value.
This SCPI is Meas Local and Context Sensitive. This SCPI alias to :CALC:MON:MARK[1]|2|3|4:MODE
POS; :CALC:MON:MARK[1]|2|3|4:X?
9.91E+37
A.13.00
:CALCulate[1]|2:MARKer[1]|2|...|4:Y?
CALC:MARK3:Y? Turn on maker 3 in Normal mode and query the Y axis value.
This SCPI is Meas Local and Context Sensitive. This SCPI alias to :CALC:MON:MARK[1]|2|3|4:MODE
POS; :CALC:MON:MARK[1]|2|3|4:Y?
A.13.00
:CALCulate[1]|2:DELTamarker[1]|2|...|4[:STATe] ON|OFF|1|0
:CALCulate[1]|2:DELTamarker[1]|2|...|4[:STATe]?
CALC:DELT3 ON Turn on maker 3 in Delta mode.
EMI Receiver Mode Reference