Example
Notes
Range à SA
This key will propagate the setting of the currently selected range to the Swept SA measurement in the
Spectrum Analyzer Mode. This functionality will enable you to use the continuous sweep capabilities that
are not available in the Frequency Scan measurement. The affected setting are:
• Start Frequency
• Stop Frequency
• RBW
• Sweep Point, which will be converted from the Step Size and Start/Stop Frequency using formula
Sweep Point = ((Stop Frequency-Start Frequency)/Step Size)+1.
• Sweep Time, which will be converted from the Dwell Time using formula Sweep Time = Dwell Time *
Sweep Point.
• Mechanical Attenuation
• Internal Preamp
• EMC Standard
• Reference Level
Key Path
Remote Command
Example
Dependencies
Couplings
Initial S/W Revision

Range Presets

Opens a menu that enables presetting of the selected scan range to CISPR or MIL bands.
EMI Receiver Mode Reference
[:SENSe]:SCAN[1]|2|...|10:INPut:TYPE?
SCAN2:INP:TYPE INPUT1
This command above is included for ESU compatibility. The suffix [1]|2 is irrelevant to the
measurement, they are added solely for remote language compatibility.
Meas Setup, Scan Table
[:SENSe]:FSCan:RANGe[1]|2|...|10:SANalyzer[:IMMediate]
FSC:RANG5:SAN Copies the parameters from range 5 of scan table to Spectrum Analyzer mode.
This key is grayed out during multi range scanning. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict; Function not available while running multirange scan" warning.
This key is grayed out during final measurement. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
The step size will be converted into sweep point when it switches into SA mode, Point = (Range
Stop- Range Start)/Step Size)+1. Point will be rounded to the nearest integer.
A.07.00
8  Frequency Scan Measurement
Meas Setup
469