Bias:state; Bias:voltage[:Level] - Keysight E4980AL User Manual

Precision lcr meter
Hide thumbs Also See for E4980AL:
Table of Contents

Advertisement

SCPI Command Reference
E4980A/AL commands
Description
Sets the DC bias range to AUTO (ON) or FIX (OFF).
When the DC bias range is FIX, "#" is displayed in the BIAS field of the display.
When the DC bias range is set to FIX, resolution of the test signal voltage is
fixed, and accuracy is changed. For details, refer to "DC bias signal, Test signal
voltage" in
Parameter
ON or 1 (Preset value)
OFF or 0
Equivalent key
No equivalent key is available on the front panel.

:BIAS:STATe

Syntax
:BIAS:STATe {ON|OFF|1|0}
:BIAS:STATe?
Description
Enables DC bias. DC bias is automatically turned off after recalling the state
from memory. When DC bias is set to ON, DC bias that has been set with
:BIAS:CURRent[:LEVel]
Parameter
ON or 1
OFF or 0 (Preset value)
Equivalent key
[DC Bias]

:BIAS:VOLTage[:LEVel]

Syntax
:BIAS:VOLTage[:LEVel] <numeric>
:BIAS:VOLTage[:LEVel]?
Description
Sets the DC bias voltage. Setting does not implicitly turn the DC bias ON. If the
bias is set by :BIAS:CURRent[:LEVel], the query command returns error No.
230. The parameter resolution depends on the setting voltage value.
328
data
sheet.
Description
Sets the DC bias range to ON
Sets the DC bias range to OFF
or
:BIAS:VOLTage[:LEVel]
Description
Enables the DC bias
Disables the DC bias
is output.