Remote Correction Data Set Commands - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

4  Input/Output Functions
Input/Output

Remote Correction Data Set Commands

This section describes the remote (SCPI) commands used to put values into correction sets. See the
correction / table editor section of the Input/Output section for the information on front panel entry of
correction data.
"Set (Replace) Data (Remote Command Only)" on page 194
"Merge Correction Data (Remote Command Only)" on page 194
Set (Replace) Data (Remote Command Only)
The command takes an ASCII series of alternating frequency and amplitude points, each value separated
by commas.
The values sent in the command will totally replace all existing correction points in the specified set.
An Ampcor array can contain 2000 points maximum.
Remote Command
Example
Preset
State Saved
Min
Max
Initial S/W Revision
Modified at S/W Revision
Merge Correction Data (Remote Command Only)
The command takes an ASCII series of alternating frequency and amplitude points, each value separated
by commas. The difference between this command and Set Data is that this merges new correction points
into an existing set.
Any new point with the same frequency as an existing correction point will replace the existing point's
amplitude with that of the new point.
An Ampcor array can contain 2000 total points, maximum.
Remote Command
Example
194
[:SENSe]:CORRection:CSET[1]|2|...|8:DATA <freq>, <ampl>, . . .
[:SENSe]:CORRection:CSET[1]|2|...|8:DATA?
CORR:CSET1:DATA 10000000, –1.0, 20000000, 1.0
This defines two correction points at (10 MHz, –1.0 dB) and (20 MHz, 1.0 dB) for correction set 1.
Empty after Restore Input/Output Defaults.  Survives a shutdown or restart of analyzer application
(including a power cycle).
Saved in instrument state.
Freq:  0 Hz
Amptd:  –1000 dBm
Freq:  1 THz
Amptd:  +1000 dBm
A.02.00
A.14.00 
[:SENSe]:CORRection:CSET[1]|2|...|8:DATA:MERGe <freq>, <ampl>, ...
CORR:CSET1:DATA:MERGE 15000000, –5.0, 25000000, 5.0
This adds two correction points at (15 MHz, –5.0 dB) and (25 MHz, 5.0 dB) to whatever values
already exist in correction set 1.