Preset
State Saved
Initial S/W Revision
Select Marker
Display a menu with 12 markers available for selection for the current measurement.
Key Path
Notes
Preset
State Saved
Initial S/W Revision
Relative To
Select the reference marker for a marker in Delta mode.
Key Path
Remote Command
Example
Notes
Preset
State Saved
Min
Max
Initial S/W Revision
Marker Trace
Assign the specified marker to the designated trace.
Key Path
Remote Command
EMI Receiver Mode Reference
Marker 1
The number of the selected marker is saved in instrument state.
A.14.00
Marker
Front panel only. The selected marker is remembered even when not in the Marker menu and is used
Marker 1
The number of the selected marker is saved in instrument state.
A.07.00
Marker, Properties
:CALCulate:DANalyzer:MARKer[1]|2|...|12:REFerence <integer>
:CALCulate:DANalyzer:MARKer[1]|2|...|12:REFerence?
CALC:DAN:MARK5:REF 1
CALC:DAN:MARK5:REF?
A marker cannot be relative to itself so that choice is grayed out, and if sent from SCPI, generates
error –221: "Settings conflict; marker cannot be relative to itself."
When queried, a single value will be returned - the specified marker number's relative marker.
2
Saved in instrument state
1
12
A.14.00
Marker, Properties
:CALCulate:DANalyzer:MARKer[1]|2|...|12:TRACe 1|2
12 Disturbance Analyzer Measurement
Marker