6  System Functions
System

Restore Align Defaults

This selection causes the Alignment system settings to be a reset to their default values. This does not
affect any Alignment data stored in the system. This level of Restore System Defaults does not affect any
other system settings, mode settings and does not cause a mode switch.
After performing this function, it may impact the auto-alignment time of the instrument until a new
alignment baseline has been established.
Confirmation is required to restore the factory default values. The confirmation dialog is:
Key Path
Example
Initial S/W Revision
Restore Misc Defaults
This selection causes miscellaneous system settings to be reset to their default values. With this reset,
you lose the GPIB address and it is reset to 18, so this should be used with caution. This level of Restore
System Defaults does not affect any other system settings, mode settings and does not cause a mode
switch. This miscellaneous group contains the rest of the settings that have not been part of the other
Restore System Defaults groups. The following table is a complete list of settings associated with this
group:
Miscellaneous Setting
Verbose SCPI
The SYST:PRES:TYPE
Auto File Name Number
Save Type
State Save To
Screen Save To
DISP:ENABle
Full Screen
342
System, Restore System Defaults
:SYST:DEF ALIG
Prior to A.02.00
Default Value
Off
MODE
000
State
Register 1
SCREEN000.png
ON
Off
EMI Receiver Mode Reference