Key Path
Remote Command
Example
Notes
Preset
State Saved
Min
Max
Default Unit
Status Bits/OPC
dependencies
Initial S/W Revision
Scale Type
Chooses a linear or logarithmic scaling for the frequency axis.
The scaling can be changed at any time and will determine only how the data will be displayed. It will have
no impact on the sweep or trace data. Changing the scaling does not restart the sweep and has no impact
on the number of sweep points. The scaling can be changed while traces are in View and they will scale
appropriately. Markers will stay at their set frequency, so they will move on the display.
Note that trace data saved while the display is in log will look identical to trace data saved while the
display is in linear. When recalling trace data, the current value of log or linear scaling will be used to
display the data. (Trace +State files will recall with whatever log or linear setting was in effect when they
were saved).
Key Path
Remote Command
Example
EMI Receiver Mode Reference
FREQ Channel
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
FREQ:CENT:STEP:AUTO ON
FREQ:CENT:STEP 500 MHz
FREQ:CENT UP increases the current center frequency value by 500 MHz
FREQ:CENT:STEP?
FREQ:CENT:STEP:AUTO?
Preset and Max values are dependent on Hardware Options (503, 508, 513, 526)
Auto
ON
Saved in instrument state
– ( the maximum frequency of the instrument). (that is, a 27 GHz max freq instrument has a CF step
range of +/– 27 GHz)
The maximum frequency of the instrument. (that is, a 27 GHz max freq instrument has a CF step
range of +/– 27 GHz)
Hz
non-overlapped
A.07.00
FREQ Channel
:DISPlay:VIEW[1]:WINDow[1]:TRACe:X:SPACing LINear | LOGarithmic
:DISPlay:VIEW[1]:WINDow[1]::TRACe:X:SPACing?
DISP:VIEW:WIND:TRAC:X:SPAC LIN
DISP:VIEW:WIND:TRAC:X:SPAC?
8 Frequency Scan Measurement
FREQ Channel