Current Measurement Query (Remote Command Only) - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

10  Monitor Spectrum Measurement
Meas
measurement and you send READ:ACP? it will initiate a new measurement using the same instrument settings.
• Initiates the measurement and puts valid data into the output buffer. If a measurement other than the current one is
specified, the instrument will switch to that measurement before it initiates the measurement and returns results.
• For example, suppose you have previously initiated the ACP measurement, but now you are running the channel
power measurement. Then you send READ:ACP? It will change from channel power back to ACP and, using the
previous ACP settings, will initiate the measurement and return results.
• Blocks other SCPI communication, waiting until the measurement is complete before returning the results
• If the optional [n] value is not included, or is set to 1, the scalar measurement results will be returned. If the [n] value
is set to a value other than 1, the selected trace data results will be returned. See each command for details of what
types of scalar results or trace data results are available. The binary data formats should be used when handling large
blocks of data since they are smaller and faster than the ASCII format. (FORMat:DATA)
Initial S/W Revision

Current Measurement Query (Remote Command Only)

This command returns the name of the measurement that is currently running.
Remote Command
Example
Initial S/W Revision
Limit Test Current Results (Remote Command Only)
Queries the status of the current measurement limit testing. It returns a 0 if the measured results pass
when compared with the current limits. It returns a 1 if the measured results fail any limit tests.
Remote Command
Example
Initial S/W Revision
Data Query (Remote Command Only)
Returns the designated measurement data for the currently selected measurement and subopcode.
n = any valid subopcode for the current measurement. See the measurement command results table for
your current measurement, for information about what data is returned for the subopcodes.
This command uses the data setting specified by the FORMat:BORDer and FORMat:DATA commands and
can return real or ASCII data. (See the format command descriptions under Input/Output in the Analyzer
Setup section.)
770
Prior to A.02.00
:CONFigure?
CONF?
Prior to A.02.00
:CALCulate:CLIMits:FAIL?
CALC:CLIM:FAIL? queries the current measurement to see if it fails the defined limits.
Returns a 0 or 1: 0 it passes, 1 it fails.
Prior to A.02.00
EMI Receiver Mode Reference