10  Monitor Spectrum Measurement
Recall
Couplings
Readback
Backwards
Compatibility SCPI
Initial S/W Revision
Amplitude Correction
These keys let you select which Correction to import the data into. Once selected, the key returns back to
the Import Data menu and the selected Correction number is annotated on the key. The next step is to
select the Open key in the Import Data menu.
Antenna corrections are a particular kind of Amplitude Corrections – they are distinguished in the
corrections file by having the Antenna Unit set to a value other than None. Only Correction 1 supports
Antenna Units.
Key Path
Notes
Dependencies
Preset
State Saved
Initial S/W Revision

Default Directory

Specifies the default directory used for loading of Amplitude Correction. If user is selected, the default
recall directory is the default (My Documents\amplitudeCorrections) or the last directory you saved the
amplitude correction data to.  Otherwise, the default recall directory is My Documents\EMC Limits and
Ampcor\Ampcor, which contains a set of preloaded amplitude correction files in the directory called
Ampcor. 
Key Path
Remote Command
Example
Notes
Preset
832
When a correction file is loaded from mass storage, it is automatically turned on (Correction ON) and
ApplyCorrections is set to On. This allows you to see its effect, thus confirming the load.
selected Correction
:MMEMory:LOAD:CORRection ANTenna | CABLe | OTHer | USER, <filename>
For backwards compatibility, ANTenna maps to 1, CABle maps to 2, OTHer maps to 3 and USER
maps to 4
A.02.00
Recall, Data, Amplitude Correction
auto return
Only Correction 1 may be used to load a Correction that contains an Antenna Unit other than None
Not part of Preset, but is reset to Correction 1 by Restore Input/Output Defaults; survives shutdown.
The current Correction number is saved in instrument state
A.02.00
Recall, Export Data, Amplitude Correction
:MMEMory:LOAD:CORRection:DDIRectory USER|PRELoaded
:MMEM:LOAD:CORR:DDIR USER Select user
:MMEM:LOAD:CORR:DDIR?
USER = User
PRELoaded = Preloaded
PRELoaded