Key Path
Remote Command
Example
Dependencies
Couplings
Preset
State Saved
Range
Readback
Initial S/W Revision
Points/ RBW
Choose the Step Size to change automatically based on number of Points/ RBW
Key Path
Example
Couplings
Initial S/W Revision
Specifies the value of Point/ RBW for the selected Range.
Key Path
Remote Command
Example
Dependencies
EMI Receiver Mode Reference
Meas Setup, Scan Table
[:SENSe]:FSCan:SCAN[1]|2|...|10:RULes PRBW|LPCT
[:SENSe]:FSCan:SCAN[1]|2|...|10:RULes?
FSC:SCAN3:RUL LPCT Set Range 3 to use auto rules of Log % of Freq
FSC:SCAN3:RUL?
This key is grayed out during multi range scanning. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict; Function not available while running multirange scan" warning.
This key is grayed out during final measurement. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
Changing Auto Step Size Rules will automatically change the step size of the selected range to Auto.
PRBW
Saved in instrument state.
Points/RBW | Log % of Freq
If Auto Step Size Rules is Points/RBW, [Points/RBW,<int>]
If Auto Step Size Rules is Log % of Freq, [Log %,<int>]
A.07.00
Meas Setup, Scan Table, Auto Step Size Rules
FSC:SCAN:RUL PRBW Selects the Points/RBW Auto Step Rules for range 1.
Selecting the points/RBW will cause the step size of the selected range to turn into Auto, even if it
was already selected.
A.07.00
Meas Setup, Scan Table, Auto Step Size Rules
[:SENSe]:FSCan:SCAN[1]|2|...|10:PRBW <integer>
[:SENSe]:FSCan:SCAN[1]|2|...|10:PRBW?
FSC:SCAN2:PRBW 2 Set 2 points/ RBW for Range 2.
FSC:SCAN2:PRBW?
This key is grayed out during multi range scanning. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict; Function not available while running multirange scan" warning.
This key is grayed out during final measurement. If the grayed out key is pressed, an advisory
8 Frequency Scan Measurement
Meas Setup
463