Test Limits (All Limits)

Selects whether displayed traces are tested against displayed limits (i.e. those for which Limit On/Off is set
to On).
For each displayed trace for which a Limit is turned on, a message will be displayed in the upper-left corner
of the graticule to notify whether the trace passes or fails the limits.
If the trace is at or within the bounds of all applicable limits and margins, the text "Trace x Pass" will be
displayed in green, where x is the trace number.  A separate line is used for each reported trace.
If the trace is at or within the bounds of all applicable limits, but outside the bounds of some applicable
margin, the text "Trace x Fail Margin" will be displayed in amber, where x is the trace number.  A separate
line is used for each reported trace.
If the trace is outside the bounds of some applicable limits, the  text "Trace x Fail" will be displayed in red,
where x is the trace number. A separate line is used for each reported trace.
If the trace has no enabled limits, or the trace itself is not displayed, no message is displayed for that trace.
The PASS/FAIL box in the corner of the Meas Bar is only displayed if there is at least one "Trace Pass/Fail"
indication displayed in the graticule.
If two amplitude values are entered for the same frequency, a single vertical line is the result. In this case,
if an upper line is chosen, the lesser amplitude is tested. If a lower line is chosen, the greater amplitude is
tested.
This command only affects the display, and has no impact on remote behavior. Limit queries over SCPI test
the trace against the limit regardless of whether the trace or the limit is turned on  (exception:  the query
:CALCulate:MONitor:TRACe[1]|2|3|4|5|6:FAIL? tests only the limits that are turned on for that trace).
Key Path
Remote Command
Example
Preset
State Saved
Range
Initial S/W Revision
Delete Limit
Deletes the currently selected limit line. Pressing Delete Limit purges the data from the limit line tables.
Limit data,  including secondary parameters such as description, margin value, etc.., will be cleared and
returned to factory preset settings.
EMI Receiver Mode Reference
Meas Setup, Limits
:CALCulate:LLINe:TEST ON|OFF|1|0
:CALCulate:LLINe:TEST?
CALC:LLIN:TEST ON
CALC:LLIN:TEST?
On, not affected by Mode Preset, preset by Restore Mode Defaults.
Saved in instrument state
On|Off
A.13.00