Alert - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

The "Align 20 Hz to 3.6 GHz required" Error Condition will appear when this alignment has expired.  It is
now your responsibility to perform the Align Now, 20 Hz to 3.6 GHz to keep the receiver in warranted
operation. This alignment can only be performed by the user as it is not part of the Auto Align process.
Key Path
Mode
Remote Command
Example
Notes
Dependencies
Couplings
Status Bits/OPC
Dependencies
Initial S/W Revision

Alert

Setting Alert to ON/OFF will enable/disable the display of RF Preselector alignment required message on
the status line.  The instrument will power up with Alert On mode.
Key Path
Mode
Remote Command
EMI Receiver Mode Reference
System, Alignments, RF Preselector, Align Now
All
:CALibration:RFPSelector:FULL
:CALibration:RFPSelector:FULL?
:CAL:RFPS:FULL
For model N9038A only.
:CALibration:RFPSelector:FULL? Return 0 if successful
:CALibration:RFPSelector:FULL? Return 1 if failed
When Align 20 Hz to 3.6 GHz is performing the alignment, bit 0 in the Status Operation register is
set. Completion, or termination, will clear bit 0 in the Status Operation register.
This command is sequential; it must complete before further SCPI commands are processed.
Interrupting the alignment from remote is
accomplished by invoking Device Clear followed by the :ABORt
command.  Successful completion will clear bit 1, bit 2 in the Status Questionable Calibration
Extended Needed register and bit 0, bit 1 in Status Questionable Calibration Extended Failure
register.
A failure encountered during alignment will set the Error Condition "20 Hz to 3.6 GHz Alignment
Failure" and set bit1, bit 2 in the Status Questionable Calibration Extended Needed register and bit 9
in Status Questionable Calibration register.
This key only appears in N9038A models, setting or querying the SCPI in other models will generate
an error.
Initializes the time for the Last Align Conducted Now, Conducted Time.
Initializes the time for the Last Align Radiated Now, Radiated Time.
Records the temperature for the Last Align Conducted Now, Conducted Temperature.
Records the temperature for the Last Align Radiated Now, Radiated Temperature.
Bit 8 or 9 may be set in the Status Questionable Calibration register.
Bit 1 and 2 may be set in the Status Questionable Calibration Extended Needed register.
Bit 0 and 1 may be set in the Status Questionable Calibration Extended Failure register.
A.08.00
System, Alignments, RF Preselector
All
:CALibration:RFPSelector:ALERt ON|OFF|0|1
:CALibration:RFPSelector:ALERt?