8 Frequency Scan Measurement
Meas Setup
State Saved
Range
Readback Text
Initial S/W Revision
Remote Command
Example
Notes
Preset
State Saved
Backwards
Compatibility SCPI
Initial S/W Revision
Time Domain Setup
Pressing this key, when it is not selected, selects Time Domain as the Scan Type. Pressing this key, when
it is already selected, accesses the Time Domain setup options. The two possible Time Domain scan types
are "Impulse" (suitable for pulsed and continuous signals) and "CW" (optimized for continuous signals).
Key Path
Remote Command
Example
Preset
State Saved
Initial S/W Revision
482
Saved in instrument state
Smooth|Discrete|Time Domain
Smooth|Discrete|Time Domain
A.07.00
[:SENSe]:FREQuency:MODE CW|FIXed|SWEep|SCAN|TDOMain
[:SENSe]:FREQuency:MODE?
FREQ:MODE SCAN
FREQ:MODE?
This command is included for ESU compatibility. The suffix [1]|2 of SENSe node is irrelevant to the
measurement, they are added solely for remote language compatibility.
CW, FIX and SWEep alias to Smooth.
SCAN aliases to Stepped. For CXA, SCAN aliases to Smooth.
TDOMain alias to Time Domain. For models that is not supporting Time Domain Scan, TDOM aliases
to Smooth.
The query will return:-
SWEep when Smooth is selected.
SCAN when Discrete is selected.
SWEep
Saved in instrument state.
[:SENSe]:FSCan:SCAN:TYPE STEPped|SWEpt|TDOMain
:SENSe1|SENSe2:FREQuency:MODE STEPped|SWEpt|TDOMain
A.07.00
Meas Setup
[:SENSe]:FSCan:SCAN:TDOMain:TYPE IMPulse|CWAve
[:SENSe]:FSCan:SCAN:TDOMain:TYPE?
FSCan:SCAN:TDOMain:TYPE IMPulse
FSCan:SCAN:TDOMain:TYPE?
IMPulse
Saved in instrument state
A.13.50
EMI Receiver Mode Reference