11  APD Measurement

AMPTD Y Scale

AMPTD Y Scale
Displays the menu keys that enable you to control the amplitude parameters.
Key Path
Notes
Initial S/W Revision
Attenuation
Enables you to set the value of the Attenuation parameter. This key only affects the Mechanical Attenuator
in the EMI Receiver mode and has no Auto setting.
For the Frequency Scan measurement, this key only affects the attenuation used for meters.
The following amplitude parameters are not settableby the user and therefore do not appear in any
menus:
Elec Attenuator
Elec Attenuation
Meas Atten Step
Max Mixer Level
Key Path
Remote Command
Example
Preset
State Saved
Min
Max
Default Unit
Backwards
Compatibility SCPI
Backwards Compatibility
SCPI Notes
Initial S/W Revision
896
Front-panel key
Front panel only.
A.13.00
Disabled
0dB
2dB
–10dBm
AMPTD Y Scale
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
[:SENSe]:POWer[:RF]:ATTenuation?
POW:ATT 10
POW:ATT?
10 dB
Saved in instrument state.
0 dB
50 dB (CXA)
60 dB (EXA)
70 dB (MXA, PXA& N9038A)
dB
:INPut[1]|2:ATTenuation
This command above is included for ESU compatibility. The suffix [1]|2 is irrelevant to the
measurement, they are added solely for remote language compatibility. This SCPI is Meas Local and