4  Input/Output Functions
Input/Output
Remote Command
Example
Notes
Notes
State Saved
Initial S/W Revision
Exit
Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired
calibration data.
Key Path
Notes
Initial S/W Revision
Q Port
The Q port calibration is performed with the front panel's Q port connected via a short BNC cable to the Cal
Out port. The guided calibration will show a diagram of the required connections.
Back
Return to the prior step in the calibration procedure.
Key Path
Notes
Initial S/W Revision
Next
Perform the Q port calibration.
212
:CALibration:IQ:FLATness:IBAR
CAL:IQ:FLAT:IBAR
The recommended procedure is to use the same BNC cable to calibrate all I/Q ports. All I/Q ports
should be calibrated sequentially during the procedure.
The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is not
reset by any preset or restore data commands.
The I-bar port must be connected to the Cal Out port before issuing the SCPI command.
No
Prior to A.02.00
Input/Output, I/Q, I/Q Cable Calibrate...
Using the Exit button will not restore the calibration data to the state prior to entering the guided
calibration. Once a port is calibrated the data is stored immediately and the only way to change it is
to redo the calibration step.
When the calibration may be left in an inconsistent state, a confirmation dialog will be displayed (see
"Exit Confirmation" on page 220
Prior to A.02.00
Input/Output, I/Q, I/Q Cable Calibrate...
Using the Back button will not restore the calibration data to a prior state. Once a port is calibrated
the data is stored immediately and the only way to change it is to redo the calibration step. The Back
button allows the user to go back to a prior step to redo that calibration step.
Prior to A.02.00
).