8  Frequency Scan Measurement
FREQ Channel
Key Path
Remote Command
Example
Notes
Dependencies
Preset
State Saved
Min
Max
Default Unit
Status Bits/OPC
dependencies
Initial S/W Revision

Start Frequency

Sets the frequency at the left side of the graticule display.
Key Path
Remote Command
Example
Notes
Dependencies
404
FREQ Channel
[:SENSe]:FREQuency:MIDSpan <freq>
[:SENSe]:FREQuency:MIDSpan?
FREQ:MIDS 140 MHZ
FREQ:MIDS?
The Max values are depending on Hardware Options (503, 507, 508, 513, 526)
TheFrequency can be limited by Start or Stop Freq limits, if the Span is so large that Start or Stop
reaches their limit.
515 MHz
Saved in instrument state.
Depends on the instrument minimum frequency and 10 Hz minimum span. If the knob or step keys
are being used, it depends on the value of the other three interdependent parameters.
Depends on the instrument maximum frequency – 10 Hz minimum span. If the knob or step keys are
being used, it depends on the value of the stop frequency
Hz
non-overlapped
A.08.00
FREQ Channel
[:SENSe]:FREQuency:STARt <frequency>
[:SENSe]:FREQuency:STARt?
[:SENSe]:FREQuency:STARt:AUTO ON|OFF|1|0
[:SENSe]:FREQuency:STARt:AUTO?
FREQ:STAR 3 GHZ
FREQ:STAR?
FREQ:STAR:AUTO 1
FREQ:STAR:AUTO?
Max values are dependent on Hardware Options (503, 507, 508, 513, 526)
By direct entry:
You cannot set the Start frequency > the Stop frequency. You cannot set the Start frequency = the
Stop frequency. You cannot select zero span by setting Start = Stop. You cannot set the Start
Frequency to a value that would create a span of less than 10 Hz. If you try to do any of these, the
Stop Frequency will change to maintain a minimum value of 10 Hz for the difference between Start
EMI Receiver Mode Reference