8  Frequency Scan Measurement
View/ Display
display area.
Key Path
Remote Command
Example
Preset
State Saved
Initial S/W Revision

Meters Freq Line

When on, a blue vertical line showing the location of the meters frequency will appear on the frequency
scan graph. In the signal list, the signal number is the same color as the Meters Freq Line when the Meters
Freq is matched with the signal's frequency. Otherwise, the signal number is in white.
Key Path
Remote Command
Example
Dependencies
Preset
State Saved
Initial S/W Revision
Next Range
This key allows you to switch across ranges on the scan trace to see the setting of the RBW, VBW,
Attenuation, Int Preamp, Dwell Time, Step Size, Scan Time and Scan Points on the graph annotation and
meas bars. The green color outline will be moved as you switch across the ranges on the trace. This key
will do nothing for a single range scan, where the green outline will not be shown.
Key Path
Notes
604
View/ Display
:DISPlay:FSCan:VIEW:METer[:STATe] ON|OFF|1|0
:DISPlay:FSCan:VIEW:METer[:STATe]?
:DISP:FSC:VIEW:MET 1
:DISP:FSC:VIEW:MET?
ON
Saved in instrument state.
A.07.00
View/ Display
:DISPlay:FSCan:VIEW:METer:FLINe ON|OFF|1|0
:DISPlay:FSCan:VIEW:METer:FLINe?
:DISP:FSC:VIEW:MET:FLIN 1
:DISP:FSC:VIEW:MET:FLIN?
The Frequency Line will only be shown if Meters Display is set to on and when a scan is not running.
ON
Saved in instrument state.
A.07.00
View/ Display
Front panel only.