Rf Input - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

Compatibility SCPI
Initial S/W Revision
Modified at S/W Revision

RF Input

Specifies using the main RF port for the current measurement
Key Path
Example
ReadBack
Initial S/W Revision
Modified at S/W Revision
RF Input 2
Specifies using the second RF port, if supported, for the current measurement. 
See
"More Information" on page 155
Key Path
Example
Couplings
ReadBack
Initial S/W Revision
More Information
In models with two inputs, the second input usually has a different maximum frequency than the first input. 
For your convenience, the actual "Max Freq" value is allowed to go slightly higher than the nominal Max
Freq for the second input, just as is the case with the first input.
EMI Receiver Mode Reference
INPut<1|2>:TYPE?
Included for R&S ESU compatibility. In the MXE, the INPUT1 parameter is aliased to RFIN and the
INPUT2 parameter is aliased to RFIN2
A.05.01
A.14.00
Input/Output, RF Input, RF Input Port
:FEED:RF:PORT RFIN
RF Input
A.05.01
A.14.00
Input/Output, RF Input, RF Input Port
:FEED:RF:PORT RFIN2
When switching from Input 1 to Input 2:
• If the Stop Freq is above the Max Freq, it is set to the Max Freq, otherwise it does not change. 
• If the Start Freq is above (Max Freq – Min Span), it is set to (Max Freq – Min Span), otherwise it
does not change.
When switching from Input 2 to Input 1, neither the Start Freq nor the Stop Freq change.
For the Swept SA measurement,  Min Span is 10 Hz.  This may vary from measurement to
measurement.
RF Input 2
A.05.01
4  Input/Output Functions
Input/Output
155

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pxa n9030aMxa n9020aExa n9010aCxa n9000aEmi n9038a

Table of Contents