Dwell Time - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

8  Frequency Scan Measurement
Marker Function
Key Path
Notes
Preset
State Saved
Initial S/W Revision

Dwell Time

Sets the dwell time for detectors
Key Path
Remote Command
Example
Notes
Dependencies
Couplings
Preset
State Saved
Min
Max
Default Unit
Initial S/W Revision
420
Meas Setup, Detectors (Measure)
Front panel only. The selected final detector is remembered even when not in the Detector Menu.
Detector 1, not affected by Mode Preset, preset by Restore Mode Defaults.
Saved in instrument state.
A.07.00
Meas Setup, Detectors (Measure)
[:SENSe]:FSCan:FINal:DETector[1]|2|3:DWELl <dwell Time>
[:SENSe]:FSCan:FINal:DETector[1]|2|3:DWELl?
FSC:FIN:DET3:DWEL 300ms  Set dwell time of detector 3 to 300 ms.
FSC:FIN:DET3:DWEL?
Not affected by a Mode Preset, preset by a Restore Mode Defaults.
The minimum Dwell Time is coupled to the detector type:-
a) When the Quasi Peak detector is selected, the minimum Dwell Time is 0.5 ms.
b) When the EMI Average/ RMS Average detector is selected, the minimum Dwell Time is 0.1 ms.
c) When the Peak/ Negative Peak detector is selected, the minimum Dwell Time is depends on the
smallest RBW among ranges turned on in the Scan Table.
RBW≤ 10 Hz
10 Hz< RBW< 100 Hz
100 Hz< RBW< 200 Hz
200 Hz < RBW<1 kHz
1 kHz < RBW<100 kHz
≥100 kHz
This key is grayed out during final measurement. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
If "Off" is selected for detector 1, this key is grayed out and shows 200 ms.
200 ms|1 s|1 s
Saved in instrument state.
1 ms|1 s|1 s
60 s
s
A.07.00
10 ms
1 ms
1 ms
0.1 ms
0.1 ms
50 us
EMI Receiver Mode Reference