4  Input/Output Functions
Input/Output
The correction to be used for each bucket is taken from the interpolated correction curve at the center of
the bucket.

Description

Sets an ASCII description field which will be stored in an exported file.  Can be displayed in the active
function area by selecting as the active function, if desired to appear in a screen capture.
Key Path
Remote Command
Example
Notes
Preset
State Saved
Initial S/W Revision
Modified at S/W Revision
Comment
Sets an ASCII comment field which will be stored in an exported file.  Can be displayed in the active
function area by selecting as the active function, if desired to appear in a screen capture.
Key Path
Remote Command
Example
Notes
Preset
State Saved
Initial S/W Revision
Modified at S/W Revision
Edit
Invokes the integrated editing facility for this correction set. 
When entering the menu, the editor window turns on, the selected correction is turned On, Apply
Corrections is set to On, the amplitude scale is set to Log, and the Amplitude Correction ("Ampcor") trace is
displayed. The actual, interpolated correction trace is shown in green for the selected correction. Note that
since the actual interpolated correction is shown, the correction trace may have some curvature to it.   This
trace represents only the correction currently being edited, rather than the total, accumulated amplitude
190
Input/Output, Corrections, Properties
[:SENSe]:CORRection:CSET[1]|2|...|8:DESCription "text"
[:SENSe]:CORRection:CSET[1]|2|...|8:DESCription?
:CORR:CSET1:DESC "11941A Antenna correction"
45 chars max; may not fit on display if max chars used
Unaffected by a Preset. Set to empty by Restore Input/Output Defaults
Saved in instrument state.
A.02.00
A.14.00 
Input/Output, Corrections, Properties
[:SENSe]:CORRection:CSET[1]|2|...|8:COMMent "text"
[:SENSe]:CORRection:CSET[1]|2|...|8:COMMent?
:CORR:CSET1:COMM "this is a comment"
60 chars max; may not fit on display if max chars used
Unaffected by Preset. Set to empty by Restore Input/Output Defaults
Saved in instrument state
A.02.00
A.14.00 