Show Alignment Statistics

Shows alignment information you can use to ensure that the instrument is operating in a specific manner.
The Show Alignment Statistics screen is where you can view time and temperature information.
Values which are displayed are only updated when the Show Alignment Statistics screen is invoked, they
are not updated while the Show Alignment Statistics screen is being displayed. The remote commands
that access this information obtain current values.
An example of the Show Alignment Statistics screen would be similar to:
A successful Align Now, RF will set the Last Align RF temperature to the current temperature, and reset
the Last Align RF time. A successful Align Now, All or Align Now, All but RF will set the Last Align Now All
temperature to the current temperature, and reset the Last Align Now All time. A successful Align Now, All
will also reset the Last Align RF items if the RF portion of the Align Now succeeded.
Key Path
Mode
EMI Receiver Mode Reference
Initializes the time for the Last Align Conducted Now, Conducted Time.
Initializes the time for the Last Align Radiated Now, Radiated Time.
Records the temperature for the Last Align Conducted Now, Conducted Temperature.
Records the temperature for the Last Align Radiated Now, Radiated Temperature.
Bit 8 or 9 may be set in the Status Questionable Calibration register.
Bit 1 and 2 may be set in the Status Questionable Calibration Extended Needed register.
Bit 0 and 1 may be set in the Status Questionable Calibration Extended Failure register.
A.13.00
System, Alignments
All
6  System Functions
System
303