Stop Frequency

Sets the frequency at the right side of the graticule display.
Key Path
Remote Command
Example
EMI Receiver Mode Reference
and Stop.
With the knob or step keys:
Cannot increment the Start Freq to a value greater than the Stop Freq – 10 Hz. If already in zero
span, you cannot increment at all, and the first decrement will be forced to at least 10 Hz.
The Start Frequency can be limited by Span limits, if the Stop Frequency is below its preset value.
This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
None
In auto, the start frequency value will couple to the start frequency of the lowest turned-on range in
the scan table. If all ranges are off, the start frequency will be set to 30 MHz.
30 MHz
ON
Saved in instrument state.
If the knob or step keys are being used, it depends on the value of the stop frequency
When the Scale Type is set to Log, the min Start Frequency will be limited to 10 Hz. If you set the
Start Frequency to a value <10 Hz during Lin Scale Type, changing the Scale Type to Log will
automatically change the start frequency to 10Hz. However, changing the Scale Type back to Lin
type will not change the frequency to the value previously set.
Depends on the instrument maximum frequency – 10 Hz minimum span. If the knob or step keys are
being used, it depends on the value of the stop frequency
Hz
non-overlapped
:SENSe1|SENSe2:FREQuency:STARt
This command above is included for ESU compatibility. The suffix 1|2 is irrelevant to the
measurement, they are added solely for remote language compatibility.
A.07.00
FREQ Channel
[:SENSe]:FREQuency:STOP <frequency>
[:SENSe]:FREQuency:STOP?
[:SENSe]:FREQuency:STOP:AUTO ON|OFF|1|0|0|1
[:SENSe]:FREQuency:STOP:AUTO?
FREQ:STOP 10 GHZ
FREQ:STOP?