Amptd Y Scale - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

10  Monitor Spectrum Measurement

AMPTD Y Scale

AMPTD Y Scale
Displays the menu keys that enable you to control the amplitude parameters.
Key Path
Initial S/W Revision
Reference Level
Specifies the amplitude represented by the topmost graticule line.
Key Path
Remote Command
Example
Preset
State Saved
Min
Max
Default Unit
Initial S/W Revision
Attenuation
Enables you to set the value of the Attenuation parameter. This key only affects the Mechanical Attenuator
in the EMI Receiver mode and has no Auto setting.
For the Frequency Scan measurement, this key only affects the attenuation used for meters.
The following amplitude parameters are not settableby the user and therefore do not appear in any
menus:
Elec Attenuator
Elec Attenuation
Meas Atten Step
Max Mixer Level
Key Path
Remote Command
730
Front-panel key
A.13.00
AMPTD/Y Scale
:DISPlay:MONitor:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RLEVel <real>
:DISPlay:MONitor:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RLEVel?
DISP:MON:VIEW:WIND:TRAC:Y:RLEV 80 dBuV
DISP:MON:VIEW:WIND:TRAC:Y:RLEV?
106.99 dBuV
Saved in instrument state
RefLevelMin = –63.01dBuV + RefLevelOffset - ExtGain.
RefLevelMax = 206.99 dBuV  + RefLevelOffset - ExtGain
Depends on the current selected Y axis unit
A.13.00
Disabled
0dB
2dB
–10dBm
AMPTD Y Scale
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
[:SENSe]:POWer[:RF]:ATTenuation?
EMI Receiver Mode Reference

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pxa n9030aMxa n9020aExa n9010aCxa n9000aEmi n9038a

Table of Contents