8  Frequency Scan Measurement
Meas Setup
When this key is pressed a prompt is placed on the screen that says "Please press Enter or OK key to
delete limit. Press ESC or Cancel to close this dialog." The deletion is only performed if you press OK or
Enter; if so, after the deletion, the informational message "Limit deleted" appears in the MSG line.
Key Path
Remote Command
Example
Backwards
Compatibility SCPI
Backwards Compatibility
SCPI Notes
Initial S/W Revision

Delete All Limits

Deletes all limit lines. Pressing Delete All Limits purges the data from all limit line tables.
All limit data will be cleared and returned to factory preset settings.
When this key is pressed a prompt is placed on the screen that says "Please press Enter or OK key to
delete all limits.  Press ESC or Cancel to close this dialog."  The deletion is only performed if you press OK
or Enter. After a deletion, the informational message "All Limits deleted" appears in the MSG line.
Key Path
Remote Command
Example
Initial S/W Revision
Limit Line Fail? (Remote Command Only)
Tests a limit line against its associated trace. Returns a 0 if the trace is within the limit and margin, a 1 if the
trace exceeds either the limit or the margin.
Note that this command only tests one limit line – other limit lines are not tested when executing this
command. To see whether a trace passed all limits, use :CALCulate:FSCan:TRACe:FAIL?.
Note this command performs the test regardless of whether the trace or the limit is turned on, on the
display.
Remote Command
Example
Backwards
Compatibility SCPI
Backwards Compatibility
498
Meas Setup, Limits
:CALCulate:FSCan:LLINe[1]|2|...|6:DELete
CALC:FSC:LLIN3:DEL  Deletes all data for limit line 3
:CALCulate[1]|2:LIMit[1]|2|...|6:DELete
This command above is included for ESU compatibility. The suffix [1]|2 is irrelevant to the
measurement, they are added solely for remote language compatibility.
A.07.00
Meas Setup, Limits
:CALCulate:FSCan:LLINe:ALL:DELete
CALC:FSC:LLIN:ALL:DEL
A.07.00
:CALCulate:FSCan:LLINe[1]|2|...|6:FAIL?
CALC:FSCan:LLIN:FAIL?  Returns a zero if limit line 1's associated trace has no failure, 1 if there is a
margin or limit failure.
:CALCulate[1]|2:LIMit[1]|2|...|6:FAIL?
This command above is included for ESU compatibility. The suffix [1]|2 is irrelevant to the