AMPTD Y Scale

Displays the menu keys that enable you to control the amplitude parameters.
Key Path
Notes
Initial S/W Revision
Ref Level
Specifies the amplitude represented by the topmost graticule line.
Key Path
Remote Command
Example
Couplings
Preset
State Saved
Min
Max
Default Unit
Initial S/W Revision
Attenuation
Enables you to set the value of the Attenuation parameter. This key only affects the Mechanical Attenuator
in the EMI Receiver mode and has no Auto setting.
For the Frequency Scan measurement, this key only affects the attenuation used for meters.
The following amplitude parameters are not settableby the user and therefore do not appear in any
menus:
Elec Attenuator
Elec Attenuation
Meas Atten Step
Max Mixer Level
EMI Receiver Mode Reference
Front-panel key
Front-panel access only.
A.07.00
AMPTD Y Scale
:DISPlay:FSCan:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RLEVel <real>
:DISPlay:FSCan:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RLEVel?
DISP:FSC:VIEW:WIND:TRAC:Y:RLEV 10dBuV
DISP:FSC:VIEW:WIND:TRAC:Y:RLEV?
If you reduce the attenuation, the instrument may have to lower the reference level to keep it below
its allowed maximum.  This allowed maximum level is specified in the "Max" row, below, along with
other variables that affect it. 
When you increase attenuation, the reference level does not change.
106.99 dBuV
Saved in instrument state.
RefLevelMin = –63.01dBuV + RefLevelOffset - ExtGain.
RefLevelMax = 206.99 dBuV  + RefLevelOffset – External Gain
Depends on the current selected Y axis unit
A.07.00
Disabled
0dB
2dB
–10dBm
8  Frequency Scan Measurement
AMPTD Y Scale
381