6  System Functions
System

Restore Align Defaults

Initializes the alignment user interface settings, not alignment data, to the factory default values. Align
Now, All must be executed if the value of the Timebase DAC results in a change.
For front panel operation, you are prompted to confirm action before setting the alignment parameters to
factory defaults:
The parameters affected are:
Parameter
Timebase DAC
Timebase DAC setting
Auto Align State
Auto Align All but RF
Auto Align Alert
Key Path
Mode
Example
Notes
Initial S/W Revision
Backup or Restore Align Data...
Opens the utility for backing-up or restoring the alignment data.
Alignment data for the instrument resides on the hard drive in a database.  Agilent uses high quality hard
drives; however it is highly recommended the alignment data be backed-up to storage outside of the
instrument.  Additionally, for customers who use multiple CPU Assemblies or multiple disk drives, the
308
Setting
Calibrated
Calibrated value
Normal (if the instrument is not operating with default
alignment data, Off otherwise)
Off
Time & Temperature
System, Alignments
All
:SYST:DEF ALIG
Alignment processing that results as the transition to Auto Alignment Normal will be executed
sequentially; thus *OPC? or *WAI will wait until the alignment processing is complete.
Prior to A.02.00