4  Input/Output Functions
Input/Output
I Ext Gain
This function affects the I channel input. However, when Q Gain in I+jQ is set to Same as I Gain, this value is
applied to both I and Q channel inputs.
Key Path
Remote Command
Example
Dependencies
Preset
State Saved
Min
Max
Readback Text
Initial S/W Revision
Q Ext Gain
This function affects the Q channel input.
Key Path
Remote Command
Example
Dependencies
Preset
State Saved
Min
182
Input/Output, External Gain
[:SENSe]:CORRection:IQ:I:GAIN <rel_ampl>
[:SENSe]:CORRection:IQ:I:GAIN?
Set the I Ext Gain to 10 dB
CORR:IQ:I:GAIN 10
Set the I Ext Gain to –10 dB (that is, a loss of 10 dB.)
CORR:IQ:I:GAIN –10
Not available unless option BBA is installed
0 dB
This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output Defaults" or "Restore
System Defaults->All"
Saved in instrument state.
–100 dB
100 dB
I Gain, <I Ext Gain> dB
Prior to A.02.00
Input/Output, External Gain
[:SENSe]:CORRection:IQ:Q:GAIN <rel_ampl>
[:SENSe]:CORRection:IQ:Q:GAIN?
Set the Q Ext Gain to 10 dB
CORR:IQ:Q:GAIN 10
Set the Q Ext Gain to –10 dB (that is, a loss of 10 dB.)
CORR:IQ:Q:GAIN –10
Not available unless option BBA is installed.
0 dB
This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output Defaults" or "Restore
System Defaults->All"
Saved in instrument state
–100 dB