Keysight X-Series Signal Analyzer
EMI Receiver Mode Reference
8 Frequency Scan Measurement
In the Frequency Scan measurement, the analyzer scans the band
using the specified Range Table settings. The scanning can be
Smooth (Swept), or Discrete (Stepped) or Time Domain. You can
select up to three traces using different detectors simultaneously.
Next, the measurement uses the Peak Excursion and Peak
Threshold values, and optionally the margin to determine which
peak to report. A maximum of 2000 peaks can be recorded for one
measurement.
For each of the peaks in the list, a final measurement will be
performed using final detectors. This measurement provides up to
three final detectors. Those with a final measurement amplitude
greater than the limit will be denoted by a red limit delta in the signal
list table.
This topic contains the following sections:
"Measurement Commands for Frequency Scan" on page 378
"Remote Command Results for Frequency Scan" on page 379
"Measurement Description" on page 380
377