Freq/ Step - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

Freq/ Step

Changes the step size for the Frequency.
Key Path
Remote Command
Example
Notes
Preset
State Saved
Min
Max
Default Unit
Status Bits/OPC
dependencies
Initial S/W Revision
EMI Receiver Mode Reference
FREQ Channel
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
FREQ:CENT:STEP:AUTO ON
FREQ:CENT:STEP 500 MHz
FREQ:CENT UP increases the current center frequency value by 500 MHz
FREQ:CENT:STEP?
FREQ:CENT:STEP:AUTO?
Preset and Max values are dependent on Hardware Options (503, 508, 513, 526)
Auto
ON
Saved in instrument state
– ( the maximum frequency of the instrument). (that is, a 27 GHz max freq instrument has a CF step
range of +/– 27 GHz)
The maximum frequency of the instrument. (that is, a 27 GHz max freq instrument has a CF step
range of +/– 27 GHz)
Hz
non-overlapped
A.07.00
9  Strip Chart Measurement