Dwell Time

Specifies the dwell time for each step size defined in the selected range.
Key Path
Remote Command
Example
Dependencies
EMI Receiver Mode Reference
CALC:FSC:SCAN:DISP OFF  Turn off scan table editor.
CALC:FSC:SCAN:DISP?
G1
Saved in instrument state.
G1|G2|OFF
A.07.00
Meas Setup, Scan Table
[:SENSe]:FSCan:SCAN[1]|2|...|10:TIME <time>
[:SENSe]:FSCan:SCAN[1]|2|...|10:TIME?
[:SENSe]:FSCan:SCAN[1]|2|...|10:TIME:AUTO ON|OFF|1|0
[:SENSe]:FSCan:SCAN[1]|2|...|10:TIME:AUTO?
FSC:SCAN7:TIME 0.5  Set the dwell time of Range 7.
FSC:SCAN7:TIME?
FSC:SCAN:TIME:AUTO ON ! Turn on Auto Dwell Time in Range 1.
FSC:SCAN:TIME:AUTO?
This key is grayed out when the Step Control is set to "Scan Time & Points". If the grayed out key is
pressed, an advisory message is generated. If the equivalent SCPI command is sent, this same
message is generated as part of a "–221, Settings conflict; Function not available while running
multirange scan" warning.
This key is grayed out during multi range scanning. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict; Function not available while running multirange scan" warning.
This key is grayed out during final measurement. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
When in Discrete (Stepped) Scan, there is no Auto setting for measurement time. The Auto/Man line
on the Dwell Time softkey disappears in this case, and if the SCPI command [:SENSe]:SCAN[1]
|2|3|4|5|6|7|8|9|10:TIME:AUTO ON is sent, it generates an error.
In Time Domain Scan, the dwell times are based on the following table:
10 Hz
100 Hz
200 Hz
1 kHz
9 kHz
100 s
100 s
100 s
100 s
100 s
8  Frequency Scan Measurement
Meas Setup
10 ms
1 ms
1 ms
100 us
100 us