Stop Freq - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

8  Frequency Scan Measurement
Meas Setup
Couplings
Preset
State Saved
Min
Max
Default Unit
Backwards
Compatibility SCPI
Backwards Compatibility
SCPI Notes
Initial S/W Revision

Stop Freq

Specifies the stop frequency of the selected range.
Key Path
Remote Command
Example
Notes
Dependencies
454
in zero span, you cannot increment at all, and the first decrement will be forced to at least 10 Hz.
If you change the start frequency of the selected range to a value < the previous range's stop
frequency (where the previous range is turned on), the stop frequency of the previous range will be
changed to the same value.
If you change the start frequency <=min frequency of the instrument, the start frequency of the
selected range will be set to the minimum frequency of the analyzer, and all the previous ranges
before the selected range will be turned off.
If you change the start frequency >=maximum frequency of the instrument, the start frequency of the
selected range will be set to the maximum frequency of the instrument –10 Hz; while the stop
frequency of selected range will be set to the maximum frequency of the instrument. All subsequent
ranges after the selected range will be turned off.
9 kHz|150 kHz|30 MHz|300 MHz|30 MHz|1 GHz|1 GHz|1.0001 GHz|1.0002 GHz|1.0003 GHz
Saved in instrument state.
If the knob or step keys are being used, depends on the value of the other three interdependent
parameters
Depends on the instrument maximum frequency – 10 Hz minimum span. If the knob or step keys are
being used, it depends on the value of the other three interdependent parameters
Hz
:SENSe1|SENSe2:SCAN[1]|2|...|10:STARt
[:SENSe]:SCAN[1]|2|...|10:STARt
This command above is included for ESU compatibility. The suffix 1|2 is irrelevant to the
measurement, they are added solely for remote language compatibility.
A.07.00
Meas Setup, Scan Table
[:SENSe]:FSCan:SCAN[1]|2|...|10:STOP <freq>
[:SENSe]:FSCan:SCAN[1]|2|...|10:STOP?
FSC:SCAN:STOP 150e3  Set stop freq of Range 1 to 150 kHz.
FSC:SCAN:STOP?
Max values depend on Hardware Options (503, 507, 508, 513, 526)
This key is grayed out during multi range scanning. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict; Function not available while running multirange scan" warning.
This key is grayed out during final measurement. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
By direct entry:
You cannot set the Stop frequency < the Start frequency. You cannot set the Start frequency = the
EMI Receiver Mode Reference