8  Frequency Scan Measurement
Meas Setup
Couplings
Preset
State Saved
Range
Readback
Initial S/W Revision

Scan Table

The scan table is used to enter the settings for up to ten ranges.
Key Path
Initial S/W Revision
Select Range
Specifies the selected range. The term "selected range" is used throughout this document to specify which
range will be affected by the functions. There are a total of 10 ranges.
Key Path
Notes
Preset
State Saved
Initial S/W Revision
Range State
Sets the state of the selected range. If a Range is turned on, it will be used as part of the measurement. If it
is off, it will be excluded. 
Key Path
Remote Command
Example
452
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict; Function not available with time domain scan type" warning.
When Time Domain is selected as the Scan Type, the Step Control is automatically set to Step Size
& Dwell Time. Changing to another Scan Type from Time Domain does not alter the existing Step
Control (i.e. it remains as Step Size and Dwell Time).
SDWell
Saved in instrument state.
Step Size & Dwell Time| Scan Time & Points
If the Step Control is Step and Dwell, will show "Step Size & Dwell Time"
If the Step Control is Scan Time/ Points, will show "Scan Time&Pts"
A.07.00
Meas Setup
A.07.00
Meas Setup, Scan Table
Front panel only. The selected range is remembered even when not in the Scan Table Menu.
Range 5.
Saved in instrument state
A.07.00
Meas Setup, Scan Table
[:SENSe]:FSCan:SCAN[1]|2|...|10:STATe ON|OFF|1|0
[:SENSe]:FSCan:SCAN[1]|2|...|10:STATe?
FSC:SCAN4:STAT 1  Turn on Range 4.
FSC:SCAN4:STAT?