Test Trace - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

Preset
State Saved
Initial S/W Revision

Test Trace

Selects the trace that you want the limit to test. 
Key Path
Remote Command
Example
Notes
Couplings
Preset
State Saved
Range
Initial S/W Revision
Description
Provides a description of up to 60 characters by which you can easily identify the limit. The descriptions will
be stored in the exported file and can be displayed in the active function area by selecting them as the
active function, if desired to be in a screen capture.
Key Path
Remote Command
Example
Preset
State Saved
Initial S/W Revision
EMI Receiver Mode Reference
Limit 1
Saved in instrument state.
A.13.00
Meas Setup, Limits, Properties
:CALCulate:APD:LLINe[1]:TRACe MEAS|MEAS2
:CALCulate:APD:LLINe[1]:TRACe?
CALC:APD:LLIN:TRAC MEAS  Applies limit 1 to Measured Trace.
When the trace is off, the trace is not tested. If the trace is on and Limit Display is on:
a.If Test Limit is Off, the "Trace Pass/Fail" indication is displayed in the upper left corner of the
graticule.
b.If Test Limit is On, the "Trace Pass/Fail" indication is displayed in the upper left corner of the
graticule, and the PASS/FAILbox is displayed in the left corner of the Meas Bar.
This matters when testing a trace or limit line for failure, via :CALC:APD:LLIN:FAIL? or
:CALC:APD:TRAC2:FAIL?
MEAS
Saved in instrument state.
Measured| Measured 2| Exceeded Limit
A.13.00
Meas Setup, Limits, Properties
:CALCulate:APD:LLINe[1]:DESCription "string"
:CALCulate:APD:LLINe[1]:DESCription?
CALC:APD:LLIN:DESC "European APD"
"" (null String), not affected by Mode Preset, preset by Restore Mode Defaults.
Saved in instrument state.
A.13.00
11  APD Measurement
Meas Setup