Status Subsystem Command Descriptions - Keysight x-series User Reference Manual

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

3
Power summary
4
Temperature summary
5
Frequency summary
8
Calibration summary
9
Integrity summary

STATus Subsystem Command Descriptions

The STATus subsystem controls the SCPI-defined instrument status reporting structures. Each status
register has a set of five commands used for querying or masking that particular register.
Numeric values for bit patterns can be entered using decimal or hexadecimal representations. (i.e. 0 to
32767 is equivalent to #H0 to #H7FFF. It is also equal to all ones, 111111111111111) See the SCPI Basics
information about using bit patterns for variable parameters.
Operation Register
"Operation Condition Query" on page 111
"Operation Enable" on page 112
"Operation Event Query" on page 112
"Operation Negative Transition" on page 112
"Operation Positive Transition" on page 113
Operation Condition Query
This query returns the decimal value of the sum of the bits in the Status Operation Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode
Remote Command
Example
Preset
Status Bits/OPC
dependencies
Initial S/W Revision
EMI Receiver Mode Reference
The instrument hardware has detected a power unleveled
condition.
The instrument is still warming up.
The instrument hardware has detected an unlocked condition or
a problem with the external frequency reference.
The instrument has detected a hardware problem while doing
the automatic internal alignment process.
The instrument has detected a questionable measurement
condition such as: bad timing, bad signal/data, timeout
problem, signal overload, or "meas uncal".
All
:STATus:OPERation:CONDition?
STAT:OPER:COND?
0
Sequential command
Prior to A.02.00
3  Programming the Analyzer