Disturbance Level (E limit)/ Probability of Time (ρ limit)
Enables you to edit the disturbance level or probability of the current row. Depending on the APD Method,
this key can be a Disturbance Level or Probability of Time.
Key Path
Notes
Min
Max
Initial S/W Revision
Key Path
Notes
Initial S/W Revision
APD Method
Selects the APD measurement method for the selected limit, either:
1. Disturbance Level: Compare a measured emission (E meas) against a limit emissions level (E limit) at a
specified probability value.
2. Probability of Time: Compare a measured probability level (ρ meas) against a limit probability (ρ limit)
level at a specified Emissions value.
Key Path
Remote Command
Example
Dependencies
Preset
State Saved
Range
Initial S/W Revision
Disturbance Level
Set the Limit APD method to Disturbance Level.
EMI Receiver Mode Reference
Meas Setup, Limits, Edit
There is no readback value on the key.
–1000 dBm
1000 dBm
A.13.00
Meas Setup, Limits, Edit
There is no readback value on the key.
A.13.00
Meas Setup, Limits, Edit
:CALCulate:APD:LLINe[1]:METHod DLEVel|PROBability
:CALCulate:APD:LLINe[1]:METHod?
CALC:APD:LLIN:METH DLEV
CALC:APD:LLIN:METH
When Disturbance Level is selected, the limit table editor will show "Disturbance Level (E limit)" key.
When Probability of Time is selected, the limit table editor will show "Prblty of Time (ρ limit)" key.
DLEVel
Saved in instrument state.
DLEVel|PROBability
A.13.00
11 APD Measurement
Meas Setup
945