3 Programming the Analyzer
STATus Subsystem
Questionable Power Negative Transition
This command determines which bits in the Questionable Power Condition register will set the
corresponding bit in the Questionable Power Event register when the condition register bit has a negative
transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Mode
Remote Command
Example
Preset
Min
Max
Status Bits/OPC
dependencies
Initial S/W Revision
Questionable Power Positive Transition
This command determines which bits in the Questionable Power Condition register will set the
corresponding bit in the Questionable Power Event register when the condition register bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Mode
Remote Command
Example
Preset
Min
Max
Status Bits/OPC
dependencies
Initial S/W Revision
Questionable Temperature Register
"Questionable Temperature Condition" on page 137
136
All
:STATus:QUEStionable:POWer:NTRansition <integer>
:STATus:QUEStionable:POWer:NTRansition?
STAT:QUES:POW:NTR 32 50 MHz Input Power became OK for Cal will be reported to the Power
Summary of the Status Questionable register.
0
0
32767
Sequential command
Prior to A.02.00
All
:STATus:QUEStionable:POWer:PTRansition <integer>
:STATus:QUEStionable:POWer:PTRansition?>
STAT:QUES:POW:PTR 32 50 MHz Input Power became too high for Cal will be reported to the
Power Summary of the Status Questionable register.
32767
0
32767
Sequential command
Prior to A.02.00