Keysight x-series User Reference Manual page 217

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

Initial S/W Revision
Back
Return to the prior step in the calibration procedure.
Key Path
Notes
Initial S/W Revision
Next
Perform the I-bar port calibration.
Key Path
Remote Command
Example
Notes
State Saved
Initial S/W Revision
Exit
Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired
calibration data.
Key Path
Notes
Initial S/W Revision
EMI Receiver Mode Reference
attached.
Prior to A.02.00
Input/Output, I/Q, I Setup, I Probe, Calibrate
Using the Back button will not restore the calibration data to a prior state. Once a port is calibrated
the data is stored immediately and the only way to change it is to redo the calibration step. The Back
button allows the user to go back to a prior step to redo that calibration step.
Prior to A.02.00
Input/Output, I/Q, I Setup, I Probe, Calibrate
:CALibration:IQ:PROBe:IBar
CAL:IQ:PROB:IB
The I-bar port must be connected to the Cal Out port before issuing the SCPI command.
The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is not
reset by any preset or restore data commands.
No
Prior to A.02.00
Input/Output, I/Q, I Setup, I Probe, Calibrate
Using the Exit button will not restore the calibration data to the state prior to entering the guided
calibration. Once a port is calibrated the data is stored immediately and the only way to change it is
to redo the calibration step.
When the calibration may be left in an inconsistent state, a confirmation dialog will be displayed (see
"Exit Confirmation" on page 220
Prior to A.02.00
).
4  Input/Output Functions
Input/Output
217