Grid Top Level (Remote Command Only)

This command defines the top grid level in the current unit for the scan display. This command is different
from Reference level where it will adjust the scale per division based on the value set for the bottom grid
level.
Remote Command
Example
Notes
Couplings
Preset
State Saved
Min
Max
Backwards
Compatibility SCPI
Backwards Compatibility
SCPI Notes
Initial S/W Revision
EMI Receiver Mode Reference
:DISPlay[:WINDow]:TRACe[1]|2|3:Y[:SCALe]:TOP <real>
:DISPlay[:WINDow]:TRACe[1]|2|3:Y[:SCALe]:TOP?
DISP:TRAC:Y:TOP 5 
DISP:TRAC:Y:TOP?
This command is included for ESU compatibility. It is not stated in the ESU user manual, but it is
used by the EMC32 application. The suffix [1]|2|3 are irrelevant to the measurement, they are added
solely for remote language compatibility.
When this value is changed, the reference level will be changed to the same value. The Scale per
division will be changed based on:
Scale/Div = ( Top Grid Level – Bottom Grid Level) / 10
The value of the bottom grid level will be recalculated based on the exact Scale/Div value.
Similar to Ref Level, the allowed range of the Grid Level setting is affected by the Ref Level Offset
Value.
106.99 dBuV
Saved in instrument state.
–63.01 dBuV
136.99 dBuV
:DISPlay:WINDow[1]|2:TRACe[1]|2|3:Y[:SCALe]:TOP
The SCPI command is not stated in the ESU user manual, but it is used by the EMC32 application.
A.07.00
8  Frequency Scan Measurement
AMPTD Y Scale
395