8 Frequency Scan Measurement
Meas Setup
Dependencies
Couplings
Preset
Min
Max
Backwards
Compatibility SCPI
Initial S/W Revision
Final Measurement (Remote Command Only)
This command performs a Final Measurement for all signals in the peak list.
Remote Command
Example
Notes
Dependencies
Initial S/W Revision
Search Method (Remote Command Only)
This command defines the search method with which maxima are searched for within a scan.
Remote Command
Example
Notes
Initial S/W Revision
508
they are added solely for remote language compatibility.
This function is not available when a scan sequence is running. If the SCPI command is sent, the
message is generated as part of a "–221, Settings conflict" warning.
When 0, all ranges will be turned off.
1
0
10
SENSe1|SENSe2:SCAN:RANGes[:COUNt]
A.07.00
INITiate[1]|2:FMEasurement
INIT:FME
This command above is included for ESU compatibility. The suffix [1]|2 is irrelevant to the
measurement, they are added solely for remote language compatibility.
This command is alias to :FSC:SEQ:REM ALL; :FSC:SEQ REM; :INIT.
This key is grayed out when a scan sequence is running. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
A.07.00
CALCulate[1]|2:PEAKsearch|PSEarch:METHod SUBRange|PEAK
CALCulate[1]|2:PEAKsearch|PSEarch:METHod
CALC:PEAK:METH SUBR Selects subrange method.
SUBRange aliases to:FSCan:SEARch:MODE SLIM; CALC:FSC:MARK:PEAK:SEAR:MODE MAX, which
actually set Search Criteria = "Subrange and Limits" and Peak Criteria = "Highest Peak".
PEAK aliases to :FSCan:SEARch:MODE PLIM; CALC:FSC:MARK:PEAK:SEAR:MODE MAX;
FSC:SEAR:PEAK:COUN:STAT OFF, which actually set Search Criteria = "Peak Criteria and Limits",
Peak Criteria = "Highest Peak" and # of Peaks State = ON.
A.11.00
EMI Receiver Mode Reference