"Questionable Temperature Enable" on page 137
"Questionable Temperature Event Query" on page 137
"Questionable Temperature Negative Transition" on page 138
"Questionable Temperature Positive Transition" on page 138

Questionable Temperature Condition

This query returns the decimal value of the sum of the bits in the Questionable Temperature Condition
register.
The data in this register is continuously updated and reflects the current conditions.
Mode
Remote Command
Example
Preset
Status Bits/OPC
dependencies
Initial S/W Revision
Questionable Temperature Enable
This command determines which bits in the Questionable Temperature Condition Register will set bits in
the Questionable Temperature Event register, which also sets the Temperature Summary bit (bit 4) in the
Questionable Register. The variable <integer> is the sum of the decimal values of the bits you want to
enable.
Mode
Remote Command
Example
Preset
Min
Max
Status Bits/OPC
dependencies
Initial S/W Revision
Questionable Temperature Event Query
This query returns the decimal value of the sum of the bits in the Questionable Temperature Event register.
EMI Receiver Mode Reference
All
:STATus:QUEStionable:TEMPerature:CONDition?
STAT:QUES:TEMP:COND?
0
Sequential command
Prior to A.02.00
All
:STATus:QUEStionable:TEMPerature:ENABle <integer>
:STATus:QUEStionable:TEMPerature:ENABle?
STAT:QUES:TEMP:ENAB 1 Reference Oscillator Oven Cold will be reported to the Temperature
Summary of the Status Questionable register.
32767
0
32767
Sequential command