8  Frequency Scan Measurement
Meas Setup
Key Path
Remote Command
Example
Dependencies
Couplings
Initial S/W Revision
Selecting a CISPR Preset will set the start/ stop frequencies, step size (according to sweep point), and
RBW of the selected range according to the table values shown below. Point/RBW will be set to 2 for
Discrete and Smooth scan type.
CISPR Band
A
9 kHz
Start Frequency
150 kHz
Stop Frequency
200 Hz
RBW
1411
Sweep Point
100 Hz
Step Size
The sweep point is based on two points for each RBW width. It is the smallest odd number ≥ 2 × (Stop
Frequency—Start Frequency)/RBW. However, in this measurement, we are using step size instead of
470
Meas Setup, Scan Table, Range Presets
[:SENSe]:FSCan:RANGe[1]|2|...|10:PRESet CISA | CISB | CISC | CISD |
CISCD | CISE | MIL1K | MIL10K | MIL150K | MIL30M | MIL1G | MILGT1G
FSC:RANG:PRES CISA  Preset Range 1 to CISPR Band A setting.
FSC:RANG:PRES CISB  Preset Range 1 to CISPR Band B setting.
FSC:RANG:PRES CISC  Preset Range 1 to CISPR Band C setting.
FSC:RANG:PRES CISD  Preset Range 1 to CISPR Band D setting.
FSC:RANG:PRES CISCD  Preset Range 1 to CISPR Band CD setting.
FSC:RANG:PRES CISE  Preset Range 1 to CISPR Band E setting.
FSC:RANG:PRES MIL1K  Preset Range 1 to MIL Std 1 kHz setting.
FSC:RANG:PRES MIL10K  Preset Range 1 to MIL Std 10 kHz setting.
FSC:RANG:PRES MIL150K  Preset Range 1 to MIL Std 150 kHz setting.
FSC:RANG:PRES MIL30M  Preset Range 1 to MIL Std 30 MHz setting.
FSC:RANG:PRES MIL1G  Preset Range 1 to MIL Std 1 GHz setting.
FSC:RANG:PRES MILGT1G  Preset Range 1 to MIL Std >1 GHz (Greater Than 1GHz) setting.
This key is grayed out during multi range scanning. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict; Function not available while running multirange scan" warning.
This key is grayed out during final measurement. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
Selecting CISPR Range Preset sets the EMI Standard in Mode Setup to CISPR. Selecting MIL
Standard Preset sets the EMI Standard to MIL. It will also set the Y Axis Unit to dBµV (unless dBµV is
grayed out, in which case it will leave the Y Axis Unit unaffected).
A.07.00
B
C
150 kHz
30 MHz
30 MHz
300 MHz
9 kHz
120 kHz
6635
4501
4.5 kHz
60 kHz
D
CD
300 MHz
30 MHz
1 GHz
1 GHz
120 kHz
120 kHz
11667
16167
60 kHz
60 kHz
E
1 GHz
Max freq of analyzer or 18
GHz, whichever is lower
1 MHz
Refer below.
500 kHz
EMI Receiver Mode Reference