Keysight x-series User Reference Manual page 728

Signal analyzer
Hide thumbs Also See for x-series:
Table of Contents

Advertisement

10  Monitor Spectrum Measurement
Dependencies
In Monitor Spectrum, there are five parameters with some interdependencies (limits and couplings).  The
diagram below shows the interdependencies:
These dependencies are detailed in the key descriptions for Spectrum RBW and Span.
The measurement time is determined by the Span as shown in the table below and is annotated in the
lower right corner of the Spectrum display.  If this time is longer than the Meas Time that is set for the
meters, the meters will use the Meas Time of the Spectrum display.
Spectrum RBW
Meas Time
10 Hz
268.5 ms
30 Hz
82.99 ms
100 Hz
26.85 ms
300 Hz
8.299 ms
1 kHz
2.685 ms
3 kHz
829.9 µs
10 kHz
268.5 µs
30 kHz
82.99 µs
100 kHz
26.85 µs