8  Frequency Scan Measurement
Meas Setup
Max
Default Unit
Initial S/W Revision

Detector

Specifies the selected detector to be used for the Final Measurement and Measure at Marker, or to turn
the selected Detector off. If you use the front panel to change the detector type, the following message
will appear to remind you that the change of the detector will cause the data of the particular detector type
column to be discarded.
Changing detector type will discard the detector values in the signal list, are you sure you want to do this?
Please press ENTER or OK to proceed. Press ESC or Cancel to close this dialog.
Key Path
Remote Command
Example
Notes
Dependencies
Couplings
Preset
State Saved
Range
Initial S/W Revision
478
60 s
s
A.07.00
Meas Setup, Detectors (Measure)
[:SENSe]:FSCan:FINal:DETector[1] | 2 | 3 POSitive | QPEak | EAVerage |
RAVerage | AVERage | NEGative | OFF
[:SENSe]:FSCan:FINal:DETector[1]|2|3?
FSC:FIN:DET3 EAV  Set detector 3 to EMI Average.
FSC:FIN:DET3?
Not affected by Mode Preset, preset by Restore Mode Defaults.
The Average and EMI detectors are mutually exclusive. If the Average detector is selected, the other
three EMI Detectors are grayed out, and vice versa. If the grayed out key is pressed, an advisory
message is generated. If the equivalent remote command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
This key is grayed out during final measurement. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a "–221, Settings conflict" warning.
If the Measure at Marker window is being displayed, and one of the detectors is changed, any value
being displayed for that detector changes to "---" until the next successful reading from that
detector.
Any change of detector type will not be reflected on the signal list, until the next Final Measurement
or Measure at Marker à List operation. Any values being displayed for that detector changes to "---
", until you perform a remeasure function for all the frequency values. 
POSitive|QPEak|EAVerage
Saved in instrument state.
Peak|Quasi Peak|EMI Average|RMS Average|Average|Negative Peak|Off
A.07.00
EMI Receiver Mode Reference