4  Input/Output Functions
Input/Output
Q Port
The Q port calibration is performed with the probe body attached to the front panel's Q port and the probe
tip connected via an adapter to the Cal Out port. The guided calibration will show a diagram of the required
connections.
Show Adapter
Show a connection diagram and instructions for the probe and adapter. See
page
220.
Key Path
Notes
Initial S/W Revision
Back
Return to the prior step in the calibration procedure.
Key Path
Initial S/W Revision
Next
Perform the Q port calibration.
Key Path
Remote Command
Example
Notes
State Saved
Initial S/W Revision
Exit
Exit the calibration procedure. All ports calibrated before pressing Exit will use the newly acquired
calibration data.
Key Path
Notes
218
Input/Output, I/Q, Q Setup, Q Probe, Calibrate
Either a passive or an active probe adapter diagram will be shown, depending on the type of probe
attached.
Prior to A.02.00
Input/Output, I/Q, Q Setup, Q Probe, Calibrate
Prior to A.02.00
Input/Output, I/Q, Q Setup, Q Probe, Calibrate
:CALibration:IQ:PROBe:Q
CAL:IQ:PROB:Q
The Q port must be connected to the Cal Out port before issuing the SCPI command.
The calibration data is saved as soon as the port is calibrated and will survive power cycles. It is not
reset by any preset or restore data commands.
No
Prior to A.02.00
Input/Output, I/Q, Q Setup, Q Probe, Calibrate
Using the Exit button will not restore the calibration data to the state prior to entering the guided
calibration. Once a port is calibrated the data is stored immediately and the only way to change it is
"Show Adapter Screen" on
EMI Receiver Mode Reference